Glossary & Definitions
Non-Destructive Testing & Consulting Services
- Absorb
dose:
The amount of energy imparted by ionizing radiation per
unit mass of irradiated matter. Indicated by "rad"; 1
rad = 0.01j/kg. SI units is "gray"; 1 gray = 1
J/kg.
See also gray (Gy), SI units
- Absorb dose rate:
the absorbed dose per unit of time; rad/s or as SI unit
gray/s.
See also gray (Gy), Kerma
- Absorption:
The process whereby the incident particles or photons
of radiation are reduced in a number or energy as they
pass through matter, i.e. the energy of the radiation
beam is attenuated. Note that the total attenuation is
the sum of the components due to photoelectric
absorption, Rayleigh scattering, Compton scattering and
pair production.
See also Attenuation
- Accelerating
potential:
The difference in electric potential between the
cathode and anode in an X-ray tube through which a
charged particle is accelerated, usually specified in
units of kV or MV.
- Activity:
The number of nuclear transitions occurring in a given
quantity of radioactive material per unit of time. For
example one disintegration/second is a becquerel (Bq),
which has replaced curie (Ci) as the standard unit of
activity.
- Acute radiation
syndrome:
The immediate effects of a short term, whole body
overexposure of a person to ionizing radiation. These
effects include nausea and vomiting, malaise, increase
temperature, and blood change.
- Ageing
fog:
The increase in optical density on an unexposed film,
measured after processing, due to long-term storage.
- Alpha particle:
The nuclei of a helium atom (with two neutrons and two
protons each) that are discharged by radioactive decay
of many heavy elements, such as uranium-238 and
plutonium-239.
- Anode:
The positive electrode of an X-ray tube. In an X-ray
tube, the anode carries the target.
- Anode
current:
The electrons passing from the cathode to the anode in
an X-ray tube. There is a small loss incurred by the
back scatted fraction.
- Artifact (false
indication):
A spurious indication on a radiograph arising e.g. from
faults in the manufacturing, handling, exposing or
processing of a film.
- Attenuation:
The reduction in intensity of a beam of X- or gamma
radiation during its passage through matter caused by
absorption and scattering.
- Attenuation
coefficient ?:
The relationship between the intensity (I0)
of a radiation incident on one side of an absorber and
the transmitted intensity (I) for an absorber thickness
(t) as expressed by I0 10 exp (-?t).
- Average gradient:
The slope of a line drawn between two specified points
on the sensitometric curve.
- Back
scatter/back scattered radiation:
Radiation which is scattered at an angle of more than
90? in relation to the direction of the incident beam.
- Beam
angle:
The angle between the central axis of the radiation
beam and the plane of the film.
- Becquerel:
The SI unit of activity equal to one disintegration per
second. [37 billion (3.7x1010) becquerels = 1 curie
(Ci)].
- Betatron:
A machine in which electrons are accelerated in a
circular orbit before being deflected onto a target to
produce high energy X-rays. This type of equipment
usually operates at energies between 10 and 31 MEV.
- Blocking medium:
A absorptive material surrounding specimens or covering
their sections used to reduce the effect of scattered
radiation on the film or on the image detector.
See also Masking
- Build-up factor:
The ratio of the intensity of the total radiation
reaching a point, to the intensity of the primary
radiation reaching the same point.
- Calibrated
density step wedge:
A piece of film having a series of different optical
densities which have been calibrated to be used as
reference densities.
- Cassette:
A rigid or flexible light-tight container for holding
radiographic recording media (film or paper) with or
without intensifying screens, during exposure.
- Cathode:
The negative electrode of an X-ray tube.
- Characteristic
curve:
A curve (of a film) showing the relationship between
the common logarithm of exposure, log K, and the
optical density, D. Also called the D-log E curve or
the H and D curve.
- Clearing time:
The time needed for the first stage of fixing of a
film, during which the cloudiness disappears.
- Collimation:
The limiting of a beam of radiation to a form of
required dimensions, by the use of diaphragms made of
absorbing material.
- Collimator:
A device made from radiation absorbent material such as
lead or tungsten, designed to limit and define the
direction and angular divergence of the radiation beam.
- Compton scatter:
A form of scattering caused by a photon of X- or gamma
radiation interacting with an electron and suffering a
reduction of energy, the scattered radiation being
emitted at an angle to the incident direction. Since
the ejected electron has a short range in most
materials, it is not considered part of the scattered
radiation. For radiation in the energy range 100 keV to
10 MeV, it is the main factor contributing to radiation
attenuation.
- Computerized
tomography (CT):
A procedure by which an image of the detail in a chosen
plane, perpendicular to the axis of the specimen, is
computed from a large number of X-ray absorption
measurements made from many directions perpendicular to
the axis. This is computerized axial tomography and
does not apply to other means of performing tomography.
- Constant potential
circuit:
An electronic configuration which is designed to apply
and maintain a substantially constant potential within
an X-ray tube.
- Continuous
spectrum:
The range of wavelengths or quantum energies generated
by an X-ray set.
- Contrast:
See "Image contrast", "Radiation contrast", "Object
contrast" and "Visual contrast".
- Contrast medium:
Any suitable substance, solid or liquid, applied to a
material being radiographed to enhance its radiation
contrast in total or in part.
- Contrast
sensitivity (thickness sensitivity):
The smallest thickness change in specimen which
produces a descendible change in optical density on a:,
radiographic (or radioscopic image, usually expressed
as a percentage of the total specimen thickness.'
- Curie
(Ci):
The unit used to describe the intensity of
radioactivity in a sample of material. The curie is
equal to 37 billion (3.7 x 1010) disintegrations per
second, which is approximately the activity of 1 gram
of radium. The Becquerel (Bq) has replaced the Ci in
the SI system. The Becquerel (Bq) is 1 disintegration
per second.
- Decay
curve:
Tftw,activity of a radiois setope;,,plotted a alInst
time, usually As a Idg/lInoar. jelationship.
- Densitometer:
A device for the measurement, of the optical density of
a radiographic film or reflective density of a
photographic print.
- Development
(of a film or paper):
chemical or physical process which converts a latent
image into a visible image.
- Diffraction mottle:
A superimposed pattern on a radiographicimtoe due to
diffraction of the incident radiation by the material
structure.
- Dose rate rreter:
An instrument for the measurement of X- or gamma
radiation dose-rate.
- Dosemeter
(dosimeter):
An instrument for measuring the accumulated dose of X-
or gamma radiation.
- Dual focus tube:
An X-ray tube with two different sizes of focus.
- Dublex
wire image quality indicator:
An image quality indicator specifically designed to
assess the overall unsharpness of a radiographic image
and composed of a series of pairs of wire elements made
of high density metal.
- Edge-blocking
material:
Material applied around a specimen or in cavities to
obtain a more uniform absorption, to reduce extraneous
scattered radiation, and to prevent local
over-exposure, e.g, fine lead shot (see also "Blocking
medium").
- Equalizing
filter (beam flattener):
A device used to equalize the intensity across the
primary X-ray beam in megavoltage radiography and so
extend the useful field size.
- Equivalent X-ray
voltage:
The voltage of a X-ray tube which produces a radiograph
most nearly equivalent to a gamma radiograph taken with
a particular gamma-ray source.
- Exposure:
The process whereby radiation is recorded on an imaging
system.
- Exposure
calculator:
A device (for example a slide rule) which may be used
to determine the exposure time required.
- Exposure chart:
A chart indicating the time for radiographic exposures
for different thicknesses of a specified material and
for a given quality of a beam radiation,
- Exposure latitude:
The range of exposures corresponding to the useful
optical density range of the emulsion.
- Exposure time:
Duration of the process of exposing a recording medium
to radiation.
- Film
base:
The support material on which the photosensitive
emulsion is coated.
- Film gradient (G):
The slope of the characteristic curve of a film at a
specified optical density D.
- Film
illuminator (viewing screen:
(viewing screen) Equipment containing a source of light
and a translucent screen used for viewing radiographs.
- Film processing:
The operations necessary to transform the latent image
on the film into a permanent, visible image, consisting
normally of developing, fixing, washing and drying a
film.
- Film system speed:
A quantitative measure of the response of a film system
to radiation, energy, for specific exposure.
conditions.
- Filter:
Uniform layer of material, u !sual higher atomic number
than the spectmen, placed between 1hp. radiation source
and the film or e purpose of Preferentially
absorbing::. the -softer radiations.
- Fixing:
The chemical removal of silver halides from a film
emulsion after development.
- Flaw sensitivity:
The minimum flaw size detectable under specified test
conditions.
- Fluorescent
intensifying, screen:
A screen consisting of a coating of phosphors which
fluoresces when exposed to X- or gamma radiation.
- Fluorometallic
intensifying screen:
A screen consisting of a metallic foil (usually lead)
coated with a material that fluoresces when exposed to
X- or gamma radiation.
- Fluoroscopy:
The production of a visible image on a fluorescent
screen by X-rays and for direct viewing of the screen.
- Focal spot size:
The dimension across the focal spot of an X-ray tube,
measured parallel to the plane of the film or the
fluorescent screen.
- Focalspot:
The X-ray emitting area on the anode of the X-ray tube,
as seen from the measuring device.
- Focus-to-film
distance (ffd):
The shortest distance from the focus of an X-ray tube
to a film set up for a radiographic exposure.
- Fog
density:
A general term used to denote the optical density of a
processed film caused by anything other than the direct
action of image - forming radiation. It can be aging
fog, chemical fog, dichroic fog, exposure fog or
inherent fog.
- Gamma Radiation:
High-energy, short wavelength electromagnetic radiation
emitted from the nucleus of an atom. Gamma rays are
very penetrating and are shielded by dense materials
such as lead. Gamma rays are similar to X rays.
- Gamma radiography:
Radiography using a gamma-ray source.
- Gamma
rays:
Electromagnetic ionizing radiation, emitted by specific
radioactive materials.
- Gamma-ray source:
Radioactive material sealed into a metal capsule.
- Gamma-ray source
container:
A container made of dense material and having a wall
thickness sufficient to produce a very great reduction
in the intensity of the radiation emitted by the
source, so as to make it safe to handle.
- Geometric
unsharpness:
Unsharpness of a radiographic image arising from the
finite size of the source of radiation. Its magnitude
also depends on the distances of sourceto-object and
object-to-film. Also called geometric blurring or
penumbra.
- Graininess:
The visual appearance of granularity.
- Granularity:
The stochastic density fluctuations in the radiograph
superimposed on the object image.
- gray
(Gy):
The SI (International System of Units) unit of
radiation dose. The unit is named for the British
physician L. Harold Gray (1905-1965), an authority on
the use of radiation in the treatment of cancer.
See also Absorb
dose
- Half
life:
The time in which half the atoms of a radioactive
substance will have disintegrated, leaving half the
original amount. Half the residue will disintegrate in
another equal period of time. The half-life values for
radioisotopes vary widely.
- Half value
thickness (HVT):
The thickness of specified material which, when
introduced into the beam of X- or gamma radiation,
reduces its intensity by a half.
- Illuminator:
An intense white light source for viewing radiographs.
- Image contrast:
See also Visual
contrast
- Image definition:
A qualitative term used to define sharpness of
delineation of image detail in a radiograph.
- Image enhancement:
Any process which increases the image definition by
improving contrast and/or definition or reducing noise.
When accomplished by computer programmes, it is
referred to as digital image processing.
- Image intensifier:
An electronic device that provides a brighter image
than that produced by the unaided action of an X-ray
beam on a flourescent screen.
- Image
quality:
That characteristic of a radiographic image defined by
the degree of detail it shows.
- Image quality
indicator (IQI):
A device used to establish a measure of the
radiographic image quality. An IQI is commonly made
using wires or steps with holes.
- Image
quality value, IQI sensitivity:
A quantitative indication of the image quality required
or achieved.
- Incident beam axis:
The axis of the beam of X-radiation defined by the
focal spot and the tube window.
- Industrial
radiology:
The use of X-rays, gamma rays, neutrons and other
penetrating radiation for nondestructive testing.
- Inherent
filtration:
The reduction of softer radiation (lower energy) of a
radiation beam by the parts of the X-ray tube, set up
or source incapsulation components, through which the
primary beam will pass.
- Inherent
unsharpness:
The blurring of a radiographic image caused by
scattered secondary radiation in the imaging medium
such as the photographic emulsion whereby these
electrons render the silver halide grains developable.
- Intensifying
factor:
The ratio of the exposure time without intensifying
screens, to that when screens are used, to obtain the
same optical density.
- Intensifying
screen:
A material used in radiographic production to converts
a part of the ionising radiation into light or
electrons and that, when in contact with a recording
medium during exposure, improves the quality of the
radiograph, or reduces the exposure time required to
produce a radiograph or both. Also see "Fluorometallic
intensifying screen", or "Fluorescent intensifying
screen".
- Kerma:
An alternative term used for absorbed dose. Kerma is
usually used to describe the transfer of energy from
photons to electrons.
See also Absorb
dose
- Latent
image:
A condition produced in an image receptor by radiation
and capable of being converted into a visible image by
film processing.
- Linear
electron accelerator (LINAC):
A device for accelerating charged particles in a
straight line, either by a steady electric field or by
a radio-frequency electric field.
- Masking:
The application of material which limits the area of
irradiation of an object to the region undergoing
radiographic examination to minimse image deterioration
due to scatter radiation.
- Metal
screen:
A screen consisting of dense metal (usually lead) that
both filters radiation and intensifies an image by
emiting electrons when exposed to X- or gamma rays.
- Microfocus
radiography:
Radiography using an X-ray tube having a small
effective focus-size of less than 0.1mm in size. Used
for geometric enlargement of the image by projection.
- Modulation
transfer function (MTF):
The ratio of the image amplitude to the object
amplitude as a function of sinusoidal frequency
variation in the object.
- Movement
unsharpness:
Blurring of a radiographic or radioscopic image as a
result of relatitve movement of the radiation source,
object or, radiation detector.
- Neutron:
An neutral hadron that is stable in the atomic nucleus
but decays into a proton, an electron and an
antineutrino with a half-life of 12 minutes outside of
the nucleus. In the nucleus it has a rest mass slightly
greater than the proton and a neutral charge.
- Object contrast:
The relative difference of radiation transmitted
between two regions of an irradiated object.
- Object-to-film
distance:
The distance between the radiation side of a test
object and the film used to radiograph the object as
measured along the central axis of the radiation beam.
- Panoramic exposure:
A radiographic set-up whereby several objects are
exposed simultaneously, or the full circumference of a
cylindrical specimen is exposed by the onmidirectional
characteristics of the radiation source.
- Penetrameter:
See also Image quality
indicator (IQI)
- Pressure mark:
A blemish on a radiograph, which may be light or dark
in appearance, depending on circumstances, and caused
by local pressure to the film.
- Primary radiation:
Radiation which travels along a straight line, without
scatter, from the source to the detector.
- Projective
magnification:
The degree of image size enlargement.
- Projective magnification technique Quality (of a
beam of radiation):
Also called Projection Microfocus radiography. A method
of radiography or radioscopy that provides an
enlargement of the image by the use of a distance
between the specimen and imaging system. (see
"Microfocus radiography").
- Quality of a
beam of radiation:
The penetrating ability of a specified form of
radiation, usually measured as a half-value thickness
of a specified material.
- rad
(rd):
An old unit of absorbed radiation dose. One rad is
equal to a dose of 0.01 joule of energy per kilogram of
mass (J/kg); one rad equals 0.01 gray or 10 milligrays.
"Rad" is an acronym for "radiation absorbed dose."
See also Absorb
dose
- Radiation contrast:
Differences in radiation intensity due to variation in
radiation opacity within an irradiated object.
- Radiation source:
Equipment used for generating X-rays or gamma rays or
other penetrating radiation sources(e.g. protons,
neutrons, Beta rays).
- Radiograph:
A visible image after processing produced by a beam of
penetrating radiation on a radiographic film or paper.
- Radiographic film:
An image storage medium consisting of a transparent
base, usually coated on both sides with a radiation
sensitive emulsion.
- Radiography:
The production of permanent visual image using
penetrating radiation through the material tested.
- Radioscopy:
The electronic production of a visual image by ionising
radiation on a radiation detector and displayed on a
monitor or similar screen.
- Radioisotope:
An unstable isotope of an element that decays by
emitting particles or gamma radiation or X-radiation.
- Rod
anode tube:
Uni-polar tubes with a long hollow anode in which the
target is situated at the extremity of a tubular anode.
These tubes generally produce a panoramic beam of
radiation.
- Scattered
Radiation:
Particulate or EM radiation that has undergone a change
in direction with or without a change in energy, during
its passage through intervening matter.
- Screen type film:
Radiographic fiIm designed to be used with fluorescent
or lead intensifying screens.
- SI
units:
Abbreviation for Syst?me International d'Unit?s. The
international system of units derived from the m.k.s.
units.
- Sievert:
The SI unit of absorbed dose equivalent (1
Joule/Kilogram or 100 rems).
- Source holder:
A container for a gamma ray source (sealed source).
- Source
size:
An indication of the radioactive intensity of a
radioactive mass typically in units of Bequerels
(formerly in Curies).
- Source-to-film
distance (sfd):
The distance between the source of radiation and the
film along the path of the beam of radiation.
- Spatial resolution:
The ability to form separable images of close objects.
- Specific activity:
The number of atoms of a radioactive substance that
disintigrate per unt time per unit mass of a
radioisotope.
- Step
wedge:
An object with specified thickness steps used to obtain
a radiograph of discrete densitiy values.
- Stereo radiography:
The production of two radiographs made using a source
shift exactluy parallel to the film plane.
- Target:
A high melting point metal on the end of the anode of
an X-ray tube on which the electron beam impinges and
from which the primary beam of X-rays is emitted
- Tube diaphragm:
A device, normally fixed to a tube shield in front of
the tube window, used to reduce scatter by limiting the
extent of the emergent X-ray beam.
- Tube
head:
One of the three main parts of an X-ray installation.
The tube head contains the tube in its shield, the
other two include the transformer and control panel.
- Tube
shield:
The metal container that supports the X-ray tube and
hold the coolant and electrical insulation fluid. It
also provides a means of reducing the leakage
radiation.
- Tube
shutter:
A device used on an X-ray tube, used to regulate the
X-ray beam, usually made of lead and remotely operated.
- Tube
voltage:
The potential difference between the anode and the
cathode of an X-ray tube, usually measured in kilovolts
(kV).
- Tube
window:
An area of relative transparency to X-rays in the X-ray
tube through which the radiation is emitted.
- Unsealed source:
Any radioactive material not encapsulated for safe
handling.
- Unsharpness:
A quantified value of image blurring. It is the total
of "geometric unsharpness", "inherent unsharpness" and
"movement unsharpness",
- Useful density
range:
The practical range of optical density on a radiograph.
Maximum density is determined by the film illuminator
and the minimum by the loss in flaw sensitivity.
- Vacuum cassette:
A light-tight container using a vacuum to hold a
radiographic recording media, film and screens, in
intimate contact during radiographic exposure.
- Viewing
mask:
A device used with a radiographic illuminator used to
exclude excessive transmitted light or to prevent light
from passing the edges of the radiograph.
- Visual contrast:
A density difference perceived visually between two
adjacent areas when viewing a radiograph.
- X-ray
film:
See also Radiographic
film
- X-ray
tube:
A device for generating X-rays by accelerating
electrons from a filament to strike a metal target
(anode).
- X-rays:
Electromagnetic radiation (photon), of shorter
wavelength than ultraviolet radiation. Produced by
bombardment of atoms by high-quntum-energy particles.
Radiation wavelength is from 10-11 to
10-9 m.
QC Laboratories, Inc. - Quality & Service
"The Leader in NDI QUALITY and SERVICE" for over Forty years.